Identity, purity and content: three different questions

Why three frequently confused terms require different analytical approaches.

01Identity

An identity test asks whether the principal signal is consistent with the expected substance. It may use mass, retention time or…

02Purity

Purity often describes the relative distribution of signals in a particular method. It depends on what the method detects and how…

03Content

Content is a quantitative determination relative to a known standard and stated basis, such as mass or dry substance. It requires…

01

Identity

An identity test asks whether the principal signal is consistent with the expected substance. It may use mass, retention time or another characteristic compared with a standard. Identity is a categorical question and does not answer how much material is present.

02

Purity

Purity often describes the relative distribution of signals in a particular method. It depends on what the method detects and how components respond. A high relative percentage does not exclude substances that are not visible to the selected detector.

03

Content

Content is a quantitative determination relative to a known standard and stated basis, such as mass or dry substance. It requires suitable calibration and accuracy control. HPLC area percentage should therefore not automatically be presented as the amount of material.

LABORATORY APPLICATION

Turning the principle into an auditable record

In a technical report, use three separate lines and avoid the broad term ‘quality’ as the only result. For each line state method, unit, standard, criterion and uncertainty. This reveals whether the data package answers all three questions or only part of them.

01Identity used as proof of quantity
02HPLC purity used as total composition
03Content without a standard or basis

PRIMARY & OFFICIAL SOURCES

Sources and further reading

  1. 01EMA / ICH Q2(R2) — Validation of analytical procedures↗
  2. 02EMA / ICH Q14 — Analytical procedure development↗

COMMON QUESTIONS

Questions specific to this topic

Can identity be correct while purity is low?

Yes. Presence of the expected substance does not exclude additional signals or other attributes outside specification.

Is area percent the same as content?

Not automatically. It is relative response within the method and needs suitable calibration for a quantitative conclusion.

Why state the reporting basis?

To make clear whether the result refers to total mass, dry substance, a specified salt or another defined basis.